0
|
1 |
# $Source: /cvs/G/DRV/SATECO/SATECO_base.template,v $
|
|
2 |
# $Date: 2008/10/28 14:01:25 $
|
|
3 |
# $Revision: 1.20 $
|
|
4 |
# $Author: wang_x1 $
|
|
5 |
|
|
6 |
|
|
7 |
# Required substitutions:
|
|
8 |
# NAME - prefix, e.g. X99SA-ES1-CRYOJET:
|
|
9 |
# UNIT - temperature unit
|
|
10 |
# PTCOLS - StreamDevice protocol file
|
|
11 |
# PORT - port name
|
|
12 |
|
|
13 |
# The Alarm Status of this record is used to trigger WATCHDOG
|
|
14 |
record(stringin, "$(NAME):VERSION")
|
|
15 |
{
|
|
16 |
field(DESC, "control version")
|
|
17 |
field(DTYP, "stream")
|
|
18 |
field(INP, "@$(PTCOLS) GET_VERSION $(PORT)")
|
|
19 |
field(SCAN, "1 second")
|
|
20 |
field(DISV, "0")
|
|
21 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
22 |
|
|
23 |
field(FLNK, "$(NAME):ONESHOT")
|
|
24 |
}
|
|
25 |
|
|
26 |
##############################################
|
|
27 |
# Controller Mode
|
|
28 |
##############################################
|
|
29 |
record(mbbi, "$(NAME):MODE_RBV")
|
|
30 |
{
|
|
31 |
field(DESC, "control mode")
|
|
32 |
field(DTYP, "stream")
|
|
33 |
field(INP, "@$(PTCOLS) GET_MODE $(PORT)")
|
|
34 |
|
|
35 |
field(ZRVL, "0")
|
|
36 |
field(ONVL, "1")
|
|
37 |
|
|
38 |
field(ZRST, "Local")
|
|
39 |
field(ONST, "Remote")
|
|
40 |
|
|
41 |
field(SCAN, "1 second")
|
|
42 |
field(DISV, "0")
|
|
43 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
44 |
}
|
|
45 |
|
|
46 |
record(mbbo, "$(NAME):MODE")
|
|
47 |
{
|
|
48 |
field(DESC, "control mode")
|
|
49 |
field(DTYP, "stream")
|
|
50 |
field(OUT, "@$(PTCOLS) SET_MODE $(PORT)")
|
|
51 |
|
|
52 |
field(ZRVL, "0")
|
|
53 |
field(ONVL, "1")
|
|
54 |
|
|
55 |
field(ZRST, "Local")
|
|
56 |
field(ONST, "Remote")
|
|
57 |
|
|
58 |
field(DISV, "0")
|
|
59 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
60 |
}
|
|
61 |
|
|
62 |
|
|
63 |
##############################################
|
|
64 |
# Connection Status Monitor
|
|
65 |
##############################################
|
|
66 |
record(mbbo, "$(NAME):CONN-STAT")
|
|
67 |
{
|
|
68 |
field(DESC, "connection status")
|
|
69 |
|
|
70 |
field(ZRVL, "0")
|
|
71 |
field(ONVL, "1")
|
|
72 |
field(TWVL, "2")
|
|
73 |
|
|
74 |
field(ZRST, "Disconnected")
|
|
75 |
field(ONST, "Connected")
|
|
76 |
field(TWST, "Connecting")
|
|
77 |
|
|
78 |
field(ZRSV, "MAJOR")
|
|
79 |
field(ONSV, "NO_ALARM")
|
|
80 |
field(TWSV, "MINOR")
|
|
81 |
}
|
|
82 |
record(bo, "$(NAME):CONN")
|
|
83 |
{
|
|
84 |
field(DESC, "turn on/off connection")
|
|
85 |
|
|
86 |
field(ZNAM, "Disconnect")
|
|
87 |
field(ONAM, "Connect")
|
|
88 |
field(ZSV, "MAJOR")
|
|
89 |
field(OSV, "NO_ALARM")
|
|
90 |
|
|
91 |
field(OUT, "$(NAME):CONN-DFAN.VAL PP")
|
|
92 |
field(FLNK, "$(NAME):CONN-CALC")
|
|
93 |
}
|
|
94 |
|
|
95 |
record(dfanout, "$(NAME):CONN-DFAN")
|
|
96 |
{
|
|
97 |
field(DESC, "fanout connect action")
|
|
98 |
field(OMSL, "supervisory")
|
|
99 |
|
|
100 |
field(OUTA, "$(NAME):ASYNIO.ENBL")
|
|
101 |
# field(OUTB, "$(NAME):ENBLSCAN.VAL")
|
|
102 |
field(OUTC, "$(NAME):WATCHDOG.VAL PP")
|
|
103 |
}
|
|
104 |
|
|
105 |
record(calcout, "$(NAME):CONN-CALC")
|
|
106 |
{
|
|
107 |
field(DESC, "connecting")
|
|
108 |
field(INPA, "$(NAME):CONN.VAL")
|
|
109 |
field(CALC, "A=1?2:0")
|
|
110 |
field(DOPT, "Use CALC")
|
|
111 |
field(OOPT, "Transition To Non-zero")
|
|
112 |
field(OUT, "$(NAME):CONN-STAT.VAL PP MS")
|
|
113 |
field(SCAN, "Passive")
|
|
114 |
}
|
|
115 |
|
|
116 |
record(calcout, "$(NAME):ONESHOT")
|
|
117 |
{
|
|
118 |
field(DESC, "watchdog trigger")
|
|
119 |
field(INPA, "$(NAME):VERSION.STAT")
|
|
120 |
field(CALC, "A=0?1:0")
|
|
121 |
field(OOPT, "When Non-zero")
|
|
122 |
field(OUT, "$(NAME):WATCHDOG.VAL PP MS")
|
|
123 |
field(SCAN, "Passive")
|
|
124 |
}
|
|
125 |
|
|
126 |
record(bo, "$(NAME):ENBLSCAN")
|
|
127 |
{
|
|
128 |
field(DESC, "enable/disable record scan")
|
|
129 |
field(DTYP, "Soft Channel")
|
|
130 |
field(SCAN, "Passive")
|
|
131 |
|
|
132 |
field(ZNAM, "Disable")
|
|
133 |
field(ONAM, "Enable")
|
|
134 |
}
|
|
135 |
|
|
136 |
record(bo, "$(NAME):WATCHDOG")
|
|
137 |
{
|
|
138 |
field(DESC, "connection indicator")
|
|
139 |
field(DTYP, "Soft Channel")
|
|
140 |
|
|
141 |
field(HIGH, "15")
|
|
142 |
field(SCAN, "Passive")
|
|
143 |
|
|
144 |
field(OUT, "$(NAME):CONN-STAT.VAL PP")
|
|
145 |
|
|
146 |
field(ZNAM, "Disconnected")
|
|
147 |
field(ONAM, "Connected")
|
|
148 |
field(ZSV, "MAJOR")
|
|
149 |
field(OSV, "NO_ALARM")
|
|
150 |
}
|
|
151 |
|
|
152 |
record(asyn, "$(NAME):ASYNIO")
|
|
153 |
{
|
|
154 |
field(DESC, "asyn port")
|
|
155 |
field(PORT, "$(PORT)")
|
|
156 |
}
|
|
157 |
|
|
158 |
record(calcout, "$(NAME):TRIG-CONN")
|
|
159 |
{
|
|
160 |
field(DESC, "trigger scan")
|
|
161 |
field(INPA, "$(NAME):WATCHDOG.VAL CPP MS")
|
|
162 |
field(CALC, "A")
|
|
163 |
field(OUT, "$(NAME):CONN.VAL PP MS")
|
|
164 |
field(OOPT, "Transition To Zero")
|
|
165 |
}
|
|
166 |
|
|
167 |
###########################################
|
|
168 |
# PID
|
|
169 |
###########################################
|
|
170 |
record(ao, "$(NAME):P")
|
|
171 |
{
|
|
172 |
field(DESC, "PID - proportion set-point")
|
|
173 |
field(DTYP, "stream")
|
|
174 |
field(OUT, "@$(PTCOLS) SET_P($(NAME)) $(PORT)")
|
|
175 |
|
|
176 |
field(DISV, "0")
|
|
177 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
178 |
field(PREC, "1")
|
|
179 |
}
|
|
180 |
|
|
181 |
record(ai, "$(NAME):P_RBV")
|
|
182 |
{
|
|
183 |
field(DESC, "PID - proportion readback")
|
|
184 |
field(DTYP, "stream")
|
|
185 |
field(INP, "@$(PTCOLS) GET_P $(PORT)")
|
|
186 |
|
|
187 |
field(SCAN, "1 second")
|
|
188 |
field(DISV, "0")
|
|
189 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
190 |
field(PREC, "1")
|
|
191 |
}
|
|
192 |
record(ao, "$(NAME):I")
|
|
193 |
{
|
|
194 |
field(DESC, "PID - integral set-point")
|
|
195 |
field(DTYP, "stream")
|
|
196 |
field(OUT, "@$(PTCOLS) SET_I($(NAME)) $(PORT)")
|
|
197 |
|
|
198 |
field(DISV, "0")
|
|
199 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
200 |
field(PREC, "1")
|
|
201 |
}
|
|
202 |
|
|
203 |
record(ai, "$(NAME):I_RBV")
|
|
204 |
{
|
|
205 |
field(DESC, "PID - integral readback")
|
|
206 |
field(DTYP, "stream")
|
|
207 |
field(INP, "@$(PTCOLS) GET_I $(PORT)")
|
|
208 |
|
|
209 |
field(SCAN, "1 second")
|
|
210 |
field(DISV, "0")
|
|
211 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
212 |
field(PREC, "1")
|
|
213 |
}
|
|
214 |
record(ao, "$(NAME):D")
|
|
215 |
{
|
|
216 |
field(DESC, "PID - derivative set-point")
|
|
217 |
field(DTYP, "stream")
|
|
218 |
field(OUT, "@$(PTCOLS) SET_D($(NAME)) $(PORT)")
|
|
219 |
|
|
220 |
field(DISV, "0")
|
|
221 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
222 |
field(PREC, "1")
|
|
223 |
}
|
|
224 |
|
|
225 |
record(ai, "$(NAME):D_RBV")
|
|
226 |
{
|
|
227 |
field(DESC, "PID - derivative readback")
|
|
228 |
field(DTYP, "stream")
|
|
229 |
field(INP, "@$(PTCOLS) GET_D $(PORT)")
|
|
230 |
|
|
231 |
field(SCAN, "1 second")
|
|
232 |
field(DISV, "0")
|
|
233 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
234 |
field(PREC, "1")
|
|
235 |
}
|
|
236 |
|
|
237 |
###########################################
|
|
238 |
# Heating Level
|
|
239 |
###########################################
|
|
240 |
record(ao, "$(NAME):HT-LVL")
|
|
241 |
{
|
|
242 |
field(DESC, "heater level set-point")
|
|
243 |
field(DTYP, "stream")
|
|
244 |
field(OUT, "@$(PTCOLS) SET_HEAT-LEVEL $(PORT)")
|
|
245 |
|
|
246 |
|
|
247 |
field(DISV, "0")
|
|
248 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
249 |
|
|
250 |
field(HOPR, "100")
|
|
251 |
field(LOPR, "0")
|
|
252 |
field(PREC, "1")
|
|
253 |
field(EGU, "%")
|
|
254 |
}
|
|
255 |
|
|
256 |
record(ai, "$(NAME):HT-LVL_RBV")
|
|
257 |
{
|
|
258 |
field(DESC, "heater level readback")
|
|
259 |
field(DTYP, "stream")
|
|
260 |
field(INP, "@$(PTCOLS) GET_HEAT-LEVEL $(PORT)")
|
|
261 |
|
|
262 |
field(SCAN, "1 second")
|
|
263 |
field(DISV, "0")
|
|
264 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
265 |
|
|
266 |
field(HOPR, "100")
|
|
267 |
field(LOPR, "0")
|
|
268 |
field(PREC, "1")
|
|
269 |
field(EGU, "%")
|
|
270 |
}
|
|
271 |
|
|
272 |
###########################################
|
|
273 |
# Temperature
|
|
274 |
###########################################
|
|
275 |
|
|
276 |
record(ai, "$(NAME):TEMP_RBV")
|
|
277 |
{
|
|
278 |
field(DESC, "temperature")
|
|
279 |
field(DTYP, "stream")
|
|
280 |
field(INP, "@$(PTCOLS) GET_TEMP $(PORT)")
|
|
281 |
field(FLNK, "$(NAME):TEMP-AVG")
|
|
282 |
|
|
283 |
field(SCAN, "1 second")
|
|
284 |
field(DISV, "0")
|
|
285 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
286 |
|
|
287 |
field(PREC, "2")
|
|
288 |
field(EGU, "$(UNIT)")
|
|
289 |
}
|
|
290 |
|
|
291 |
record(ao, "$(NAME):TEMP")
|
|
292 |
{
|
|
293 |
field(DESC, "temperature")
|
|
294 |
field(DTYP, "stream")
|
|
295 |
field(OUT, "@$(PTCOLS) SET_TEMP $(PORT)")
|
|
296 |
|
|
297 |
field(DISV, "0")
|
|
298 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
299 |
|
|
300 |
field(PREC, "2")
|
|
301 |
field(EGU, "$(UNIT)")
|
|
302 |
|
|
303 |
field(FLNK, "$(NAME):GO-1 PP MS")
|
|
304 |
}
|
|
305 |
|
|
306 |
#!
|
|
307 |
record(ai, "$(NAME):TEMP-SP")
|
|
308 |
{
|
|
309 |
field(DESC, "temperature setpoint")
|
|
310 |
field(DTYP, "stream")
|
|
311 |
field(INP, "@$(PTCOLS) GET_TEMP-SP $(PORT)")
|
|
312 |
|
|
313 |
field(DISV, "0")
|
|
314 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
315 |
|
|
316 |
field(PREC, "2")
|
|
317 |
field(EGU, "$(UNIT)")
|
|
318 |
}
|
|
319 |
|
|
320 |
###########################################
|
|
321 |
# busy record for temperature set-point
|
|
322 |
###########################################
|
|
323 |
record(calcout, "$(NAME):GO-1")
|
|
324 |
{
|
|
325 |
field(DESC, "")
|
|
326 |
field(INPA, "$(NAME):TEMP.SIMM")
|
|
327 |
field(CALC, "A=1?0:1")
|
|
328 |
|
|
329 |
field(OUT, "$(NAME):GO-2 PP MS")
|
|
330 |
field(OOPT, "When Non-zero")
|
|
331 |
}
|
|
332 |
|
|
333 |
record(bo, "$(NAME):GO-2")
|
|
334 |
{
|
|
335 |
field(DESC, "")
|
|
336 |
field(DTYP, "Soft Channel")
|
|
337 |
field(OMSL, "supervisory")
|
|
338 |
field(VAL, "1")
|
|
339 |
field(OUT, "$(NAME):BUSY.VAL PP MS")
|
|
340 |
# update TEMP-STAT, not waiting for next readback
|
|
341 |
field(FLNK, "$(NAME):RNG-CALC")
|
|
342 |
}
|
|
343 |
|
|
344 |
record(busy, "$(NAME):BUSY")
|
|
345 |
{
|
|
346 |
}
|
|
347 |
|
|
348 |
###########################################
|
|
349 |
# Temperature Status Monitor
|
|
350 |
###########################################
|
|
351 |
|
|
352 |
record(longout, "$(NAME):SETTL-PRD")
|
|
353 |
{
|
|
354 |
field(DESC, "settling time")
|
|
355 |
field(EGU, "s")
|
|
356 |
field(VAL, "120")
|
|
357 |
field(FLNK, "$(NAME):TEMP-AVG")
|
|
358 |
}
|
|
359 |
|
|
360 |
record(ao, "$(NAME):TOLERANCE")
|
|
361 |
{
|
|
362 |
field(DESC, "tolerance band width")
|
|
363 |
field(EGU, "$(UNIT)")
|
|
364 |
field(VAL, "1")
|
|
365 |
field(PREC, "2")
|
|
366 |
}
|
|
367 |
|
|
368 |
record(sub, "$(NAME):TEMP-AVG")
|
|
369 |
{
|
|
370 |
field(DESC, "temperature average")
|
|
371 |
field(INPA, "$(NAME):SETTL-PRD.VAL")
|
|
372 |
field(INPB, "$(NAME):TEMP_RBV.VAL")
|
|
373 |
field(INPD, "0") # Continuous Mode
|
|
374 |
field(INAM, "initSubAve")
|
|
375 |
field(SNAM, "SubAve")
|
|
376 |
field(PREC, "2")
|
|
377 |
field(SCAN, "Passive")
|
|
378 |
|
|
379 |
field(FLNK, "$(NAME):RNG-CALC")
|
|
380 |
}
|
|
381 |
|
|
382 |
record(calcout, "$(NAME):RNG-CALC")
|
|
383 |
{
|
|
384 |
field(DESC, "temperature status calc")
|
|
385 |
field(INPA, "$(NAME):TEMP-AVG.VAL")
|
|
386 |
field(INPB, "$(NAME):TEMP.VAL")
|
|
387 |
field(INPC, "$(NAME):TOLERANCE.VAL")
|
|
388 |
field(INPD, "$(NAME):TEMP_RBV.VAL")
|
|
389 |
|
|
390 |
field(CALC, "(ABS(A-B)<=C)&&(ABS(D-B)<=C)?0:1")
|
|
391 |
field(DOPT, "Use CALC")
|
|
392 |
field(OOPT, "When Zero")
|
|
393 |
|
|
394 |
field(OUT, "$(NAME):BUSY.VAL PP MS")
|
|
395 |
field(FLNK, "$(NAME):TEMP-STAT.VAL PP MS")
|
|
396 |
|
|
397 |
field(SCAN, "Passive")
|
|
398 |
}
|
|
399 |
|
|
400 |
record(mbbo, "$(NAME):TEMP-STAT")
|
|
401 |
{
|
|
402 |
field(DESC, "temperature status")
|
|
403 |
|
|
404 |
field(DOL, "$(NAME):RNG-CALC MS")
|
|
405 |
field(OMSL, "closed_loop")
|
|
406 |
|
|
407 |
field(ZRVL, "0")
|
|
408 |
field(ONVL, "1")
|
|
409 |
|
|
410 |
field(ZRST, "Inside Range")
|
|
411 |
field(ONST, "Out of Range")
|
|
412 |
|
|
413 |
field(ZRSV, "NO_ALARM")
|
|
414 |
field(ONSV, "MAJOR")
|
|
415 |
|
|
416 |
field(SCAN, "Passive")
|
|
417 |
}
|
|
418 |
###########################################
|
|
419 |
# For Debug Use
|
|
420 |
###########################################
|
|
421 |
record(stringout, "$(NAME):QUERY")
|
|
422 |
{
|
|
423 |
field(DESC, "debug - query")
|
|
424 |
field(DTYP, "stream")
|
|
425 |
field(OUT, "@$(PTCOLS) QUERY($(NAME):RECV.VAL) $(PORT)")
|
|
426 |
field(DISV, "0")
|
|
427 |
field(SDIS, "$(NAME):WATCHDOG.VAL")
|
|
428 |
}
|
|
429 |
record(stringout, "$(NAME):RECV")
|
|
430 |
{
|
|
431 |
field(DESC, "debug - received message")
|
|
432 |
field(DTYP, "Soft Channel")
|
|
433 |
}
|
|
434 |
|
|
435 |
###########################################
|
|
436 |
# retrieve temp set-point when connected
|
|
437 |
###########################################
|
|
438 |
record(calcout, "$(NAME):TRIG-STAT")
|
|
439 |
{
|
|
440 |
field(DESC, "trigger TRIG-SP")
|
|
441 |
field(INPA, "$(NAME):CONN-STAT.VAL CPP MS")
|
|
442 |
field(CALC, "A=1?1:0")
|
|
443 |
field(OUT, "$(NAME):TEMP-SIMM1.VAL PP MS")
|
|
444 |
field(OOPT, "Transition To Non-zero")
|
|
445 |
}
|
|
446 |
record(bo, "$(NAME):TEMP-SIMM1")
|
|
447 |
{
|
|
448 |
field(DESC, "set TEMP in simulation")
|
|
449 |
field(OUT, "$(NAME):TEMP.SIMM")
|
|
450 |
field(FLNK, "$(NAME):SEQ")
|
|
451 |
}
|
|
452 |
|
|
453 |
record(sseq, "$(NAME):SEQ")
|
|
454 |
{
|
|
455 |
field(SELM, "All")
|
|
456 |
|
|
457 |
field(WAIT1,"Wait")
|
|
458 |
field(DO1, "1")
|
|
459 |
field(LNK1, "$(NAME):TEMP-SP.PROC CA")
|
|
460 |
|
|
461 |
field(DOL2, "$(NAME):TEMP-SP.VAL")
|
|
462 |
field(LNK2, "$(NAME):TEMP.VAL PP")
|
|
463 |
|
|
464 |
field(FLNK, "$(NAME):TEMP-SIMM2")
|
|
465 |
}
|
|
466 |
|
|
467 |
record(bo, "$(NAME):TEMP-SIMM2")
|
|
468 |
{
|
|
469 |
field(DESC, "reset TEMP out simulation")
|
|
470 |
field(DOL, "0")
|
|
471 |
field(OUT, "$(NAME):TEMP.SIMM")
|
|
472 |
field(OMSL, "closed_loop")
|
|
473 |
}
|